HPLC荧光检测法测定人血浆中缬沙坦的浓度

赵荣伟, 俞佳   

  1. 浙江省人民医院药剂科, 杭州 310014, 浙江
  • 收稿日期:2005-06-22 修回日期:2005-07-28 发布日期:2020-11-22
  • 作者简介:赵荣伟,本科,主管药师,从事医院药学研究。Tel:(0)13357180957 E-mail:zrwzjh@sohu.com

Determination of valsartan in human plasma by HPLC with fluorimetric detection

ZHAO Rong-wei, YU Jia   

  1. Department of Pharmacy, Zhejiang Provincial Hospital, Hangzhou 310014, Zhejiang, China
  • Received:2005-06-22 Revised:2005-07-28 Published:2020-11-22

摘要: 目的: 建立HPLC 法测定人血浆中缬沙坦浓度的方法。方法: 以厄贝沙坦为内标, 血浆用液-液萃取处理。采用Agilent ZORBAX SB-C18 (150 mm ×4.6 mm, 5 μm);柱温:室温;流动相:乙腈:水:磷酸:三乙胺为40:60:1.0:1.5(v/v);流速1.0ml·min-1;荧光检测激发波长为265 nm, 发射波长为378 nm。结果: 血浆内源性杂质对样品测定无干扰。缬沙坦最低定量浓度为25 ng·ml-1, 在25 ~2 500 ng·ml-1 范围内线性关系良好(r =0.9996)。缬沙坦的日内、日间RSD 均小于8%, 样品多次冻融及提取后24 h 内稳定性良好。结论: 该法快速、灵敏、准确, 可用于缬沙坦的体内分析及临床药学研究。

关键词: 缬沙坦, HPLC, 荧光检测法, 血浆

Abstract: AIM: To establish a high performance liquid chromatography (HPLC) method to determine the concentration of valsartan in human plasma.METHODS: Using lrbesartan as internal standard, valsartan in plasma samples was determined by HPLC with liquid-liquid extraction, achieved by the column of Agilent ZORBAXSB-C18(150 mm ×4.6 mm, 5 μm) at room temperature.The mobile phase consisted of a mixture acetonitrile:water:phosphoric acid:triethylamine was the ratio of 40:60:1.0:1.5 (v/v), pumped at a flow rate of 1.0 ml·min-1, the wavelengths of fluorimetric excitation and emission were set at 265 and 378 nm respectively.RESULTS: The drug-free plasma did not interfere with the determination of drugs and internal standard.There was good linear relationships (1/C 2 weighted) between peak area ratio of valsartan to internal standard and C (r =0.9996) within the range of 25-2 500 ng·ml-1.The precision of within-day and between-day was good.The lower limit of quantification was 25 ng·ml-1.The analytes reconstituted in the mobile phase were also stable at ambient conditions for at least 24 h.Furthermore, valsartan was stable for at least three freeze thaw cycles.CONCLUSION: The HPLC method can be used to determine the concentration of valsartan in human plasma.

Key words: valsartan, HPLC, fluorescence, plasma