Abstract: AIM: To compare two methods of HPLC-MS/MS and MEIA to monitor the tacrolimus concentrations for organ transplantation patient. METHODS: After developing and validating the HPLC-MS/MS method to determine the concentrations of tacrolimus, the tacrolimus samples were quantitated by two methods of HPLC-MS/MS and MEIA, respectively.Evaluate two methods by statistical analysis of quantitative results.RESULTS: The mean concentration were(4.86±0.46)ng/mL by HPLC-MS/MS and (5.52±0.43)ng/mL by MEIA.The coefficient correlation of two methods was 0.8771, and the two methods have good correlations in statistics.The mean concentration ratio was (90.3±5.3)%.CONCLUSION: The HPLC-MS/MS method is a more precise method for to determining the effective concentration of tacrolimus, which is suitable for daily TDM.

Key words: HPLC-MS/MS, microparticle enzyme immunoassay, organ transplantation, tacrolimus