A research on application of Actiwatch to monitor the locomotor activity in mice

LIU Jing, XU Jiang-tao   

  1. Urumqi General Hospital of Lanzhou Command of PLA, Department of Neurology, Urumqi 830000, Xinjiang, China
  • Received:2011-06-16 Revised:2011-07-01 Online:2011-07-26 Published:2011-09-22

Abstract: AIM: Using Actiwatch to monitor the locomotor activity in mice and verifying the correlation between Actiwatch and manual counting.METHODS: Thirty mice were randomly divided into three groups(control, diazepam and caffeine group), compared the synchronized data of locomotor activity of mice in jiggle-cage, which recorded by manual counting and Actiwatch, then made a correlation analysis.RESULTS: Both the Actiwatch and manual counting results showed that diazepam group had significant difference from control group and caffeine group, P<0.01; However control group had little difference from caffeine group, P>0.05. The correlation coefficient of movement time and activity number which recorded by Actiwatch and manual counting respectively were 0.83 and 0.91.CONCLUSION: The data recorded by Actiwatch and manual counting have a close correlation, and Actiwatch can be used to monitor the locomotor activity in mice.

Key words: Actiwatch, Mice, Locomotor activity, Jiggle-cage